Abstract
Tetraalkylammonium salts [R(3)R'N](+)X(-) (R = R', X: CH3, Cl; C2H5, B r; C3H7, Br; C4H9, H2PO4; C5H11, Br; R, R', X: CH3, C16H33, Br) were i nvestigated as ion-pairing reagents for the reversed-phase ion-pair ch romatographic separation of selenite and selenate. Other chromatograph ic parameters such as composition and pH of the mobile phase and conce ntration of the ion-pairing reagent were also investigated. The compat ibility of the proposed chromatographic procedures with various seleni um-specific detectors is discussed. The absolute detection limits were found to be 31 ng Se for selenite and 51 ng Se for selenate (100 mu l injection) with a solution of 5 mM tetrabutylammonium dihydrogen phos phate in 50:50 (v/v) methanol/water as mobile phase at a flow rate of 1 ml/min when a flame atomic absorption spectrometer was used as detec tor. The HPLC/FAAS system was employed for the determination of seleni te in solutions serving as selenium supplement for animals.