CHARACTERIZATION OF AN IMAGE-PLATE DETECTOR USED FOR QUANTITATIVE SMALL-ANGLE-SCATTERING STUDIES

Citation
F. Ne et al., CHARACTERIZATION OF AN IMAGE-PLATE DETECTOR USED FOR QUANTITATIVE SMALL-ANGLE-SCATTERING STUDIES, Journal of applied crystallography, 26, 1993, pp. 763-773
Citations number
24
Categorie Soggetti
Crystallography
ISSN journal
00218898
Volume
26
Year of publication
1993
Part
6
Pages
763 - 773
Database
ISI
SICI code
0021-8898(1993)26:<763:COAIDU>2.0.ZU;2-M
Abstract
This paper describes the characterization of an image-plate readout sy stem used to calculate small-angle-scattering intensities as cross sec tions per unit volume. The raw data are the latent phosphorescent exci tations stored on an image plate. Self-decay, self exposure, resolutio n, smearing, dynamic range and sensitivity are measured. Preliminary e xperiments show that an inexpensive neutron-imaging system can be obta ined when the plates are placed behind a suitable neutron-to-X-ray con verter.