G. Ferrari et al., MOCK CIRCULATORY-SYSTEM FOR IN-VITRO REPRODUCTION OF THE LEFT-VENTRICLE, THE ARTERIAL TREE AND THEIR INTERACTION WITH A LEFT-VENTRICULAR ASSIST DEVICE, Journal of medical engineering & technology, 18(3), 1994, pp. 87-95
A modular physical circuit for testing monoventricular and biventricul
ar heart assist devices (HAD) is under development. The modules note,
available are assembled in an open-loop circuit and reproduce the func
tion of the left ventricle and the systemic arterial tree. The left ve
ntricle model reproduces Starling's law of the heart and can be easily
controlled to modify other parameters such as contractility and timin
g (i.e. heart rate and systole/diastole ratio). This circuit, in conne
ction with a left ventricular assist device (LVAD), can be used to eva
luate the LVAD performance, its effect on the circulatory system and a
s a training system. This paper is devoted to a description of the cir
cuit and of its interaction with a LVAD, which is analysed after the s
imulation of a low contractility pathology of the ventricle. Results o
btained in these experiments are reported.