M. Reitzerbergaentzle et al., HPLC DETERMINATION OF VITAMIN-B6 IN FOODS AFTER PRECOLUMN DERIVATIZATION OF FREE AND PHOSPHORYLATED VITAMERS INTO PYRIDOXOL, Food chemistry, 48(3), 1993, pp. 321-324
A method to determine the content of vitamin B-6 in foods by ion pair
high performance liquid chromatography is proposed and includes pre-co
lumn transformation of phosphorylated and free vitamers into pyridoxol
followed by fluorescence detection. Dephosphorylation of vitamers is
first achieved by enzymatic hydrolysis (acid phosphatase). By reaction
with glyoxylic acid in presence of Fe2+ as catalyst, pyridoxamine is
transformed into pyridoxal, which is then reduced to pyridoxol by the
action of sodium borohydride in alkaline medium. The proposed method h
as a good recovery rate (90-95%), a satisfactory repeatability (coeffi
cient of variation less than 8%) and a very low detection limit (0.02
mug g-1).