Dual-energy x-ray absorptiometry (DXA) was evaluated as a method for m
easuring the body composition of growing broiler chickens. A total of
130 chickens, ranging in weight from 400 to 3,290 g, were scanned usin
g a DXA instrument (Lunar(TM) DPX-L). Single whole-body scans were acq
uired and analyzed using pediatric total body research software (neona
tal mode) or small animal total body research software (detail or high
resolution mode). The DXA measurements provided readings of total tis
sue mass, percentage fat, fat tissue mass, lean tissue mass, and bone
mineral content. After scanning, the bodies of the chickens were froze
n, then, after removing the feathers, homogenized for chemical determi
nation of fat, water, and protein content. By chemical analysis, the w
hole body fat content of the chickens ranged from 2.8 to 27.2%, giving
rise to DXA R values (ratio of attenuation coefficients) ranging from
1.415 to 1.339. The accuracy of DXA for measuring total body fat was
a function of the scanning program and mode and also the size of the b
ird. The best agreement between DXA and chemical measurements of perce
ntage body fat were obtained when chickens weighing more than 2,000 g
were scanned using either the small animal-detail mode or neonatal mod
e. None of the scan modes proved to be accurate for measuring the fat
content of birds weighing less than 2,000 g. The DXA measurement of le
an mass of chickens was found to be highly correlated with both total
body protein (R-2 = 0.90) and total body water (R-2 = 0.93), but was o
f Little value for predicting percentage values for either. The ratio
of DXA bone mineral content to total body ash was 0.77; however, the c
orrelation (R-2) between the two was only 0.46. These results suggest
that although the DXA technique is potentially useful for measuring bo
dy composition of chickens, considerable refinement is needed prior to
routine application.