APPLICATION OF DETECTOR NONLINEARITY CORRECTION FOR FT-IR SPECTROMETRIC GAS-ANALYSIS OF SF6 SAMPLES

Citation
Hm. Heise et al., APPLICATION OF DETECTOR NONLINEARITY CORRECTION FOR FT-IR SPECTROMETRIC GAS-ANALYSIS OF SF6 SAMPLES, Mikrochimica acta, 1997, pp. 301-303
Citations number
4
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00263672
Year of publication
1997
Supplement
14
Pages
301 - 303
Database
ISI
SICI code
0026-3672(1997):<301:AODNCF>2.0.ZU;2-7
Abstract
Fourier-transform infrared spectroscopy can achieve the low spectral n oise level necessary for trace analysis. In the midinfrared a photocon ductive HgCdTe detector is routinely used, which suffers From a substa ntial non-linear response for high radiation flux such as found when r ecording the interferogram centre-burst. The photometric accuracy, clo se to detector cut-off, is thereby dramatically reduced for strongly a bsorbing samples, and the line shapes can be seriously influenced by p hase errors. The effects on the spectra of gaseous SF6, and the improv ements gained by software correction are presented.