DETERMINATION OF THE COLLECTION EFFICIENCY FOR COMBINED MAGNETIC-ELECTROSTATIC SEM OBJECTIVE LENSES

Authors
Citation
G. Knell et E. Plies, DETERMINATION OF THE COLLECTION EFFICIENCY FOR COMBINED MAGNETIC-ELECTROSTATIC SEM OBJECTIVE LENSES, Optik, 108(1), 1998, pp. 37-42
Citations number
12
Categorie Soggetti
Optics
Journal title
OptikACNP
ISSN journal
00304026
Volume
108
Issue
1
Year of publication
1998
Pages
37 - 42
Database
ISI
SICI code
0030-4026(1998)108:1<37:DOTCEF>2.0.ZU;2-E
Abstract
In this paper the calculation of the collection efficiency for a combi ned magnetic-electrostatic objective lens is presented. We have assume d that a rotationally symmetric, coaxially arranged detector is mounte d above the lens (through-lens detection). Designing a lens, one shoul d ensure that the image planes of secondary electrons are not located too close to the detector plane, because in these cases very high dete ction losses would occur. The comparison of two lens variants, which h ave only slight differences with regard to primary electron optics, sh ows large differences in collection characteristics in the case of a p rimary electron energy of 1 keV. The proportion of backscattered elect rons on the detector signal is discussed.