The FIM Group of the University of Rouen is developing a new type of d
etector that can be used in atom probe and 3D atom probe applications.
This detector consists of an array of conductive and transparent stri
ps covered with a phosphorescent material. After having produced light
, electrons generated by ionic impacts onto microchannel plates produc
e signals on strips that are used for timing. At the same time, light
produced by ions on the phosphor screen is recorded by means of a CCD
image sensor. Then, the comparison between positions and the distribut
ion of time of flight on the strip array makes it possible to correlat
e positions and times for each event. In this contribution, preliminar
y results obtained with a 5 x 5 cm(2) detector having eight strips wil
l be shown. Potential capabilities and performance will be discussed.
(C) 1998 Elsevier Science B.V. All rights reserved.